Research brief
Reading a Peptide Mass Spectrum: MALDI-TOF vs ESI
Short answer
Reading a peptide mass spectrum means turning peaks plotted on a mass-to-charge axis into a neutral molecular mass, then comparing that number with the mass calculated from an amino acid sequence. The two ionization methods a researcher is most likely to meet, matrix-assisted laser desorption/ionization coupled to a time-of-flight analyzer (MALDI-TOF) and electrospray ionization (ESI), present that information in very…
Reading a peptide mass spectrum means turning peaks plotted on a mass-to-charge axis into a neutral molecular mass, then comparing that number with the mass calculated from an amino acid sequence. The two ionization methods a researcher is most likely to meet, matrix-assisted laser desorption/ionization coupled to a time-of-flight analyzer (MALDI-TOF) and electrospray ionization (ESI), present that information in very different shapes. MALDI-TOF usually gives one dominant singly charged peak per species, so the neutral mass is one subtraction away. Electrospray spreads the same molecule across a series of multiply charged peaks, so the neutral mass has to be recovered by assigning charge states and deconvoluting the envelope. The arithmetic behind both is short and the same in principle; what differs is the number of peaks involved and the kinds of artifacts that sit beside them.
What is a peptide mass spectrum actually plotting?
A peptide mass spectrum plots detected ion abundance on the vertical axis against mass-to-charge ratio on the horizontal axis. Nothing on the plot is a molecule; every feature is an ion that survived transfer from the condensed phase into the gas phase, traversed the analyzer and reached the detector. Neutral species are invisible, which is why a compound that resists ionization can be present in a sample and absent from the spectrum.
The vertical axis is normally relative intensity, scaled so that the largest peak in the displayed range, the base peak, sits at 100 percent. That scaling is a display convention, not a measure of how much material is present. Ionization efficiency varies by sequence, by charge-carrying residue content, by solvent chemistry and, in MALDI, by how the analyte sits in the matrix crystal. Two peptides at the same concentration can differ substantially in peak height, so intensity is read as a qualitative signal unless an internal standard or a defined quantitative workflow has been built around it.
The horizontal axis is where the chemistry lives. Peaks cluster into isotope groups, charge states, adduct series and, in MALDI, matrix-derived background. Reviews of the field, including Mass spectrometry-based proteomics in Nature in 2003, describe the discipline as the systematic conversion of these ion signals into molecular information, with the caveat that interpretation depends as much on knowing the instrument's artifacts as on the arithmetic. Reading a spectrum therefore begins with identifying which features belong to the analyte and which belong to the method.
Why is the horizontal axis m/z rather than mass?
The axis is m/z because mass analyzers separate ions by the ratio of mass to charge, never by mass alone. A time-of-flight tube sorts ions by flight time, which depends on the kinetic energy an ion of charge z acquires in the accelerating field. A quadrupole transmits ions whose m/z satisfies stable trajectories at a given radiofrequency and direct-current setting. An orbital trap or an ion cyclotron resonance cell measures a frequency that scales with m/z. In each case charge is baked into the measurement, so the raw axis is a ratio and the mass has to be extracted.
The quantity z is the net number of charges on the ion. For peptides analyzed in positive mode, charge almost always comes from added protons, so an ion carrying z protons appears at m/z equal to the neutral monoisotopic mass plus z times the proton mass, all divided by z. The proton mass used in this arithmetic is approximately 1.00728, which is the mass of a hydrogen atom minus the mass of an electron. Using 1.008 instead introduces a small systematic error that matters when a measurement is being judged at parts-per-million accuracy on a high-resolution instrument and is negligible on a low-resolution one.
Because the axis is a ratio, a large molecule carrying many charges can appear at a low m/z value. This is the single most important fact in reading electrospray data: a peak at m/z 1000 says nothing about molecular size until its charge is known. In MALDI the same arithmetic applies, but because z is usually 1 the ratio and the mass nearly coincide, which is why the two methods feel so different to read.
How does MALDI-TOF mass spectrometry ionize peptides?
MALDI generates ions by firing a pulsed laser at a solid sample in which the analyte is embedded in a large excess of a small, strongly absorbing organic matrix. The matrix absorbs the laser energy, ablates into a plume and carries analyte molecules with it, where proton transfer produces mostly singly charged species. The approach entered the literature when Laser desorption ionization of proteins with molecular masses exceeding 10,000 daltons appeared in Analytical Chemistry in 1988, reporting that intact large biomolecules could be desorbed and detected when mixed with an absorbing matrix. That work was carried out on model proteins with ultraviolet laser desorption, and the mechanism it proposed has been refined since.
Mechanistic work such as Ion formation in MALDI: the cluster ionization mechanism, published in Chemical Reviews in 2003, describes ion formation as proceeding through charged clusters released from the ablated matrix, with subsequent neutralization and proton transfer determining which species survive as observable ions. The practical consequence for anyone reading a spectrum is that MALDI ion yields are heterogeneous across a crystal surface, so spectra from different positions on the same spot can differ in intensity while agreeing on m/z.
Matrix choice shapes the spectrum. Alpha-cyano-4-hydroxycinnamic acid is the matrix most often described for peptides, while sinapinic acid and dihydroxybenzoic acid appear in protein and modified-peptide work. Every matrix contributes its own ions, clusters and fragments, which crowd the low mass region and are usually the reason the displayed range starts several hundred mass units above zero. Those matrix signals are background, not analyte, and the first step in reading a MALDI spectrum is recognizing them as such.
How is a singly charged MALDI-TOF peak converted to a molecular mass?
A singly charged protonated peptide ion appears at m/z equal to the neutral monoisotopic mass plus approximately 1.00728, so the neutral mass is recovered by subtracting that proton mass from the observed value. If a peak sits at m/z 1619.82 and is confirmed as the protonated species, the neutral monoisotopic mass is about 1618.81. That single subtraction is the whole of the arithmetic, which is why MALDI-TOF mass spectrometry of peptides is often the fastest way to read an intact mass.
Confirming that the peak really is the protonated form is the part that takes judgment. Three checks are standard. First, the isotope spacing: a singly charged peptide ion shows isotope peaks separated by approximately 1 mass unit, because each additional heavy isotope, mostly carbon-13, adds roughly one unit to a single-charge ion. Spacing of about 0.5 units means the ion carries two charges, which MALDI does produce for some sequences and under some matrix conditions. Second, the presence of sodium and potassium adducts displaced from the protonated peak by about 21.98 and 37.96 units respectively, which confirm that the parent peak is the proton adduct rather than an alkali adduct of something lighter. Third, the presence of a weak signal at roughly twice the m/z value, the singly charged dimer, and at roughly half, the doubly charged ion, which together bracket the assignment.
MALDI spectra are also read with attention to the mass region immediately below the main peak, where losses of water, ammonia or a labile modification can appear. Those losses can arise in the plume rather than in solution, so a small loss peak is not automatically evidence about the material itself.
How does electrospray ionization create a multiply charged peptide ion series?
Electrospray produces a multiply charged peptide ion series because the analyte leaves solution already carrying several protons, and the population of ions distributes across a range of charge states rather than collapsing onto one. A high voltage applied between a fine emitter tip and the instrument entrance disperses an acidic aqueous-organic solution into charged droplets; as solvent evaporates the droplets shrink until gas-phase ions are released. The method was introduced to large biomolecules in Electrospray ionization for mass spectrometry of large biomolecules, published in Science in 1989, which reported that proteins and other large molecules appeared as coherent series of multiply charged ions whose spacing could be used to recover the molecular mass.
The mechanism of release has been examined in detail. Unraveling the mechanism of electrospray ionization, in Analytical Chemistry in 2013, reviews the competing descriptions, with small ions escaping by field emission from the droplet surface, globular species released as the droplet evaporates to their size, and disordered chains extruded from the droplet while charges are acquired along the chain. That last picture explains why unfolded or flexible peptides tend toward higher charge states than compact ones of similar mass, and why charge-state distributions shift when solution conditions change.
Source geometry also matters. Analytical properties of the nanoelectrospray ion source, published in Analytical Chemistry in 1996, described operation at very low solution flow with fine emitters, reporting improved signal stability and sample economy relative to conventional flow rates. The work was performed on peptide and protein test samples under defined laboratory conditions, so the reported gains describe that configuration rather than every electrospray source.
How is the charge state of an electrospray peak assigned?
Charge state is assigned either from isotope spacing within a single peak cluster or from the spacing between adjacent peaks in the charge envelope. The isotope route is the more direct when resolution allows it: adjacent isotope peaks of an ion carrying z charges are separated by approximately 1 divided by z on the m/z axis. Spacing near 1 indicates a singly charged ion, near 0.5 a doubly charged ion, near 0.33 a triply charged ion, and so on. This requires an analyzer that resolves isotopes at that m/z, which rules the method out on low-resolution instruments at higher charge.
The envelope route uses two neighbouring charge states and needs no isotopic resolution at all. If the lower m/z peak of the pair is written as m1 and the higher as m2, the charge of the higher-m/z peak follows from dividing the quantity m1 minus 1.00728 by the difference m2 minus m1. Taking a worked example, a spectrum showing peaks at m/z 1000.51 and 1500.26 gives a difference of 499.75 and a numerator of 999.50, so the charge of the 1500.26 peak is 2 and the charge of the 1000.51 peak is 3.
Both routes are checked against the rest of the envelope. Once a charge is assigned to one peak, the predicted positions of the neighbouring charge states follow immediately, and all of them should be present with a smooth intensity profile. A peak that refuses to join a series is a candidate for a different component, a salt cluster or a contaminant, and should be treated separately rather than forced into the envelope.
How is a charge-state envelope deconvoluted into one neutral mass?
Deconvolution converts each charged peak into a neutral mass estimate and then combines those estimates into one value. For a peak at m/z value m carrying charge z, the neutral mass is z multiplied by the quantity m minus 1.00728. Applying that to the worked pair above, the 2+ peak at 1500.26 gives 2 times 1499.25, or 2998.51, and the 3+ peak at 1000.51 gives 3 times 999.50, or 2998.51. Agreement between charge states is the internal consistency check that the assignment is right; disagreement means a charge has been misassigned or the peaks come from different species.
In practice software handles this across the whole envelope. Algorithms in common use either transform the spectrum onto a neutral mass axis by summing evidence from all charge states, or fit a model of expected isotope and charge patterns to the data and report the masses that best explain the observed peaks. The output is a zero-charge spectrum in which each component appears once, which is far easier to read than the raw multiply charged data.
Deconvolution also introduces its own failure modes. Harmonic artifacts can appear at half or twice the true mass when a charge series is incomplete. Overlapping envelopes from co-eluting species can be merged into one false mass. Noise at low abundance can be promoted into spurious components if thresholds are set permissively. For that reason the raw m/z spectrum is retained alongside the deconvoluted result, and any reported mass is traced back to the specific peaks that produced it.
What is the difference between monoisotopic and average mass in a spectrum?
Monoisotopic mass is calculated using only the lightest stable isotope of each element, while average mass uses the natural abundance weighted atomic masses, and the two differ by an amount that grows with molecular size. For carbon the monoisotopic value is exactly 12 by definition and the average atomic weight is slightly higher because of natural carbon-13; across a peptide containing many carbon atoms those fractions accumulate.
Which value a spectrum reports depends on resolution. When isotopes are resolved, the first peak of the cluster is the monoisotopic peak, containing only light isotopes, and that is the value compared against a monoisotopic calculation. When isotopes are not resolved, the analyzer delivers a single broad peak whose centroid approximates the average mass, and comparing that centroid against a monoisotopic calculation produces an apparent error that has nothing to do with the sample.
For small peptides the monoisotopic peak is usually the most intense peak in the cluster. As mass increases, the probability that a molecule contains at least one heavy isotope rises, the cluster broadens, and the monoisotopic peak falls in relative height until it becomes a shoulder and then disappears into the noise. At that point the monoisotopic value cannot be read directly and is either modelled from the cluster shape or abandoned in favour of the average mass. Mismatching the two conventions, calculating monoisotopic and measuring average or the reverse, is one of the most common sources of a mass that looks wrong by a few units when nothing is actually wrong.
How is an observed mass compared with a theoretical mass?
The comparison is made by calculating the expected mass from the sequence using the same convention as the measurement, then expressing the difference as an absolute value in mass units or as a relative error in parts per million. Relative error is the difference divided by the theoretical mass, multiplied by one million. At m/z 1000, an error of 5 parts per million corresponds to 0.005 mass units; at m/z 2000 the same relative error corresponds to 0.010.
The theoretical mass for a linear peptide is the sum of the residue masses plus the mass of water, since each peptide bond formation removes one water molecule. Modifications are then added or subtracted as defined mass shifts. Terminal chemistry matters: a C-terminal amide instead of a free acid lowers the mass by approximately 0.984, and an N-terminal acetyl group raises it by approximately 42.011. Cyclic and disulfide-containing structures carry their own offsets, discussed below.
What tolerance counts as acceptable depends entirely on the instrument and the context, and no single universal figure exists in the literature; it is a parameter set by the laboratory against the demonstrated accuracy of the analyzer, its calibration state and the resolution of the data. A low-resolution linear time-of-flight measurement and a high-resolution orbital trap measurement of the same peptide justify very different windows. Reporting an error in parts per million without also reporting the instrument, the calibration approach and whether the value is monoisotopic or average leaves the number uninterpretable.
Which adducts, salts and matrix signals appear beside the peptide peak?
The peaks sitting next to an analyte signal are usually alkali metal adducts, solvent or buffer adducts, matrix-derived ions, or in-source fragments. Recognizing them prevents a reader from treating an artifact as a component. Sodium adducts appear displaced from the protonated species by approximately 21.98 units in a singly charged ion, and potassium adducts by approximately 37.96; in multiply charged electrospray data the same displacements appear divided by the charge, which compresses them and can make them look like isotope features on a low-resolution instrument.
Trifluoroacetic acid, which is common in peptide chromatography solvents, contributes its own adduct and also suppresses electrospray signal, which is why formic acid is frequently described as the preferred acid for electrospray work in the analytical literature. Ammonium, phosphate and other buffer components can produce series of clusters that march across the spectrum at regular intervals; a repeating constant spacing that does not correspond to an amino acid residue or an isotope is the signature of a salt cluster series.
MALDI adds matrix background. Matrix molecules form protonated species, clusters and photochemical products that dominate below roughly m/z 500 and can extend higher. Running a matrix-only control and overlaying it on the sample spectrum is the standard way to separate matrix background from analyte signal.
In-source fragmentation, sometimes called prompt or metastable fragmentation in time-of-flight work, produces peaks below the parent that correspond to losses of water, ammonia, or labile groups such as phosphate or glycan units. These fragments are generated inside the instrument, so their presence describes the gas-phase stability of the ion rather than the composition of the sample.
How do resolving power and mass accuracy limit what can be read?
Resolving power determines whether isotopes and closely spaced species can be separated, and mass accuracy determines how closely the centre of a peak matches the true value; they are independent properties and both constrain interpretation. Resolving power is commonly expressed as the m/z value divided by the peak width at half maximum height. A spectrum that cannot resolve the isotope cluster of a 3+ ion cannot use isotope spacing to assign charge, and it cannot report a monoisotopic mass.
Mass accuracy depends on calibration, on the stability of the analyzer and, in trapping instruments, on the number of ions in the trap, since space-charge effects shift measured frequencies. Calibration is internal when reference ions of known mass are present in the same spectrum and external when the instrument is calibrated separately before the run; internal calibration is generally described as the more accurate because it corrects for drift at the moment of measurement.
Time-of-flight instruments gain resolution from reflectron optics, which compensate for the spread of initial kinetic energies, and from delayed extraction, which allows the ion packet to equilibrate before acceleration. Linear mode, with no reflectron, trades resolution for transmission and is used mainly for larger species. The practical reading consequence is simple: the mode of operation has to be known before the spectrum is interpreted, because a linear-mode peak and a reflectron-mode peak carry different information even for the same sample.
Which mass shifts correspond to common peptide modifications?
Common modifications appear as defined offsets from the calculated backbone mass, and most interpretation of unexpected masses starts by testing those offsets. Oxidation of methionine adds approximately 15.995, which is the mass of one oxygen atom. Deamidation of asparagine or glutamine adds approximately 0.984, the replacement of an amide nitrogen with oxygen, which is detectable only on an instrument that resolves and measures to better than a mass unit. Formation of a disulfide bond between two cysteine residues removes approximately 2.016, two hydrogen atoms, relative to the reduced form. Loss of ammonia from an N-terminal glutamine to form pyroglutamate removes approximately 17.027.
Acetylation adds approximately 42.011 and is distinguished from trimethylation, which adds approximately 42.047, only on instruments with sufficient accuracy to separate those two values. Phosphorylation adds approximately 79.966 and is often accompanied in fragmentation spectra by a characteristic neutral loss. Amidation of the C-terminus, which is a design feature of several synthetic research peptides, lowers the mass by approximately 0.984 relative to the free acid, a shift that is identical in size to deamidation and opposite in sign.
Two cautions apply when reading these shifts. Several shifts are near-coincident at low resolution, so a modification assignment made from an intact mass alone is provisional. And a shift observed in the spectrum does not establish where in the sequence it occurred; localization requires fragmentation data. The literature does not provide a universal table that resolves all isobaric ambiguities from intact mass alone, which is why tandem experiments exist.
How do the mass analyzers behind MALDI and ESI differ?
MALDI is most often paired with time-of-flight analyzers and electrospray with quadrupoles, ion traps, orbital traps and ion cyclotron resonance cells, because the ion sources differ in whether they deliver ions in pulses or continuously. A pulsed laser suits a time-of-flight tube, which needs a defined start time for every ion packet. A continuous spray suits scanning and trapping analyzers, which sample from an ongoing ion stream. These pairings are conventions rather than rules, and hybrid configurations exist in both directions.
High-resolution analyzers changed what can be read from peptide spectra. Fourier transform ion cyclotron resonance mass spectrometry: a primer, published in Mass Spectrometry Reviews in 1998, sets out how measuring ion cyclotron frequencies in a strong magnetic field yields very high resolving power and mass accuracy, along with the trade-offs in acquisition time and space-charge sensitivity. Electrostatic axially harmonic orbital trapping: a high-performance technique of mass analysis, in Analytical Chemistry in 2000, introduced the orbital trap, in which ions oscillate along the axis of a shaped electrode and the axial frequency is measured to give mass. A later review, Orbitrap mass spectrometry, in Analytical Chemistry in 2013, summarizes how the technique developed into routine high-resolution instrumentation and notes that resolving power on these instruments depends on transient acquisition length, so resolution and acquisition speed are traded against one another.
For a reader of spectra, the analyzer determines three things: whether isotopes are resolved at the charge state in question, how tight a mass tolerance is defensible, and which artifacts to expect. None of these can be inferred from the ionization method alone.
When is MALDI-TOF the better instrument and when is electrospray?
Neither is better in general; they answer different questions and tolerate different samples. MALDI-TOF is typically chosen when a fast intact mass is wanted from a simple sample, when the spectrum should be easy to read because charge states collapse to one, when tolerance to salts and buffers matters, and when a large number of spots are to be surveyed quickly. The singly charged output means the m/z axis is nearly a mass axis, which is why the method is so widely used for straightforward peptide mass readout.
Electrospray is typically chosen when the sample arrives from liquid chromatography, when quantitative comparisons across a run are planned, when the extra charge is useful because it brings large species into the accessible m/z range, and when fragmentation of selected precursor ions is required. Multiply charged ions also fragment more informatively in many activation methods than singly charged ones, which is one reason electrospray dominates sequencing workflows.
The trade-offs run the other way too. Electrospray is more sensitive to salts and detergents, which suppress signal and generate adduct series. MALDI spectra vary across the sample spot and are more prone to in-source fragmentation of labile groups in some matrices. Electrospray spectra require deconvolution, which adds an interpretive step that can go wrong. MALDI spectra of mixtures can hide components whose ionization is suppressed by more efficiently ionizing neighbours.
Many laboratories run both, using one as a fast readout and the other when separation, fragmentation or higher accuracy is needed.
How does tandem mass spectrometry extend reading past the intact mass?
Tandem mass spectrometry isolates one precursor ion, fragments it and records the fragment masses, so the spectrum reports sequence-related information rather than a single molecular mass. Fragment ions are named by a convention set out in Proposal for a common nomenclature for sequence ions in mass spectra of peptides, published in Biomedical Mass Spectrometry in 1984, which assigns a, b and c labels to fragments retaining the N-terminus and x, y and z labels to those retaining the C-terminus. That nomenclature is still the standard vocabulary for describing fragment series.
The ABC's (and XYZ's) of peptide sequencing, in Nature Reviews Molecular Cell Biology in 2004, reviews how fragment series are generated and read, describing collision-induced dissociation as producing mainly b and y ions and electron-based methods as producing mainly c and z ions that preserve labile modifications. Reading a tandem spectrum means finding a ladder of peaks whose spacings equal residue masses and walking that ladder in both directions.
Two computational approaches dominate interpretation. Error-tolerant identification of peptides in sequence databases by peptide sequence tags, in Analytical Chemistry in 1994, described extracting a short partial sequence from a fragment ladder and using it with flanking masses to search a database, an approach built to tolerate unexpected modifications. Probability-based protein identification by searching sequence databases using mass spectrometry data, in Electrophoresis in 1999, introduced a probability-based scoring framework for matching spectra to database entries. Both were developed for database-resident sequences, so neither applies directly to a sequence absent from the database being searched.
What are the most common errors in reading a peptide spectrum?
The most common errors are charge misassignment, isotope misassignment, convention mismatch and treating artifacts as components. Charge misassignment produces masses that are exact multiples or fractions of the truth, which is why cross-checking two charge states is worth the thirty seconds it takes. Isotope misassignment happens when the monoisotopic peak is weak or buried in noise and a heavier isotope peak is picked instead, producing an error of one or two units that looks like a chemical difference.
Convention mismatch covers comparing a monoisotopic measurement against an average calculation, forgetting the proton mass, using 1.008 where 1.00728 is required at high accuracy, or ignoring whether the calculated sequence assumed a free acid or an amidated C-terminus. These produce small, systematic, reproducible errors that can persist across an entire data set.
Artifacts misread as components include salt cluster series, matrix background, in-source fragments, dimers, and deconvolution harmonics. A simple discipline covers most of them: before reporting any mass, locate every major peak in the spectrum and give it an explanation, even if the explanation is background.
Finally, calibration drift and detector saturation are underappreciated. A saturated peak is flat-topped or distorted and its centroid is unreliable, so the mass should be re-measured at lower signal. Spectra acquired long after the last calibration should be treated as approximate until recalibrated. None of these checks require special software; they require that the raw spectrum, not only the processed report, is actually looked at.
How should a spectrum be recorded and reported in a laboratory record?
A usable record states the ionization method, the analyzer and its operating mode, the calibration approach, the acquisition range, whether the reported mass is monoisotopic or average, the observed value, the calculated value and the difference expressed in both absolute and relative terms. Without those fields a reported mass cannot be re-evaluated later by anyone else.
Raw files are retained alongside processed output so that charge assignments and deconvolution results can be re-derived. Where deconvolution software was used, the record notes the program and the settings, since thresholds and mass ranges change which components are reported. Where a modification is proposed from a mass shift, the record states whether fragmentation data supports the localization or whether the assignment rests on the intact mass alone.
Analytical data of this kind also appears in supplier documentation. Real Peptides supplies research-use-only compounds and publishes independent third-party certificates of analysis for every batch at /coas/, and its catalog of research compounds is listed at /shop/, with company location details published at /location/. A researcher reading a published analytical report applies the same questions used on an in-house spectrum: which ionization method, which analyzer, which mass convention, and what was the difference between observed and calculated values.
The habits that make a spectrum readable are the same habits that make it reportable. Identify every peak, assign charge before mass, check the assignment against a second charge state or the isotope spacing, match conventions between measurement and calculation, and keep the raw data. Those steps are independent of whether the ions were desorbed from a matrix crystal or sprayed from a solution.
All compounds discussed on this page are sold for laboratory research use only and are not for human consumption.
References
- Fenn JB, Mann M, Meng CK, Wong SF, Whitehouse CM. Electrospray ionization for mass spectrometry of large biomolecules. Science. 1989. PMID: 2675315
- Karas M, Hillenkamp F. Laser desorption ionization of proteins with molecular masses exceeding 10,000 daltons. Analytical Chemistry. 1988. PMID: 3239801
- Karas M, Krüger R. Ion formation in MALDI: the cluster ionization mechanism. Chemical Reviews. 2003. PMID: 12580637
- Konermann L, Ahadi E, Rodriguez AD, Vahidi S. Unraveling the mechanism of electrospray ionization. Analytical Chemistry. 2013. PMID: 23134552
- Wilm M, Mann M. Analytical properties of the nanoelectrospray ion source. Analytical Chemistry. 1996. PMID: 8779426
- Makarov A. Electrostatic axially harmonic orbital trapping: a high-performance technique of mass analysis. Analytical Chemistry. 2000. PMID: 10740853
- Zubarev RA, Makarov A. Orbitrap mass spectrometry. Analytical Chemistry. 2013. PMID: 23590404
- Marshall AG, Hendrickson CL, Jackson GS. Fourier transform ion cyclotron resonance mass spectrometry: a primer. Mass Spectrometry Reviews. 1998. PMID: 9768511
- Roepstorff P, Fohlman J. Proposal for a common nomenclature for sequence ions in mass spectra of peptides. Biomedical Mass Spectrometry. 1984. PMID: 6525415
- Steen H, Mann M. The ABC's (and XYZ's) of peptide sequencing. Nature Reviews Molecular Cell Biology. 2004. PMID: 15340378
- Mann M, Wilm M. Error-tolerant identification of peptides in sequence databases by peptide sequence tags. Analytical Chemistry. 1994. PMID: 7847635
- Perkins DN, Pappin DJ, Creasy DM, Cottrell JS. Probability-based protein identification by searching sequence databases using mass spectrometry data. Electrophoresis. 1999. PMID: 10612281
- Aebersold R, Mann M. Mass spectrometry-based proteomics. Nature. 2003. PMID: 12634793