Research brief
Reversed-Phase HPLC for Peptides: Columns, Gradients, 214 nm
Short answer
Reversed-phase HPLC is a liquid chromatography mode in which the stationary phase is non-polar, usually an alkyl chain bonded to porous silica, and the mobile phase is polar, usually water mixed with acetonitrile and a small amount of acid. In reversed-phase HPLC, peptides adsorb to that hydrophobic surface while the eluent is mostly aqueous, then release when the organic fraction…
Reversed-phase HPLC is a liquid chromatography mode in which the stationary phase is non-polar, usually an alkyl chain bonded to porous silica, and the mobile phase is polar, usually water mixed with acetonitrile and a small amount of acid. In reversed-phase HPLC, peptides adsorb to that hydrophobic surface while the eluent is mostly aqueous, then release when the organic fraction climbs past a composition specific to each sequence, which is why the technique can resolve chains that differ by a single residue. The standard configuration for peptide purity work is a C18 column, a water and acetonitrile gradient acidified with trifluoroacetic acid, and ultraviolet detection at 214 nm, where the amide bond of the peptide backbone absorbs. The sections below walk through each part of that method, what controls retention and selectivity, and where the information on a chromatogram stops.
What is reverse phase HPLC, and why is it the default mode for peptide analysis?
Reverse phase HPLC is high performance liquid chromatography in which the stationary phase is less polar than the mobile phase, the inverse of the older normal phase arrangement where a polar silica surface was eluted with non-polar organic solvents. In the reversed arrangement, retention rises as the analyte becomes more hydrophobic and falls as the organic content of the eluent rises. A pump delivers the mobile phase through a packed column at high pressure, a sample is carried onto the head of the column in a mostly aqueous eluent, and a detector downstream records what leaves the column as a function of time.
The mode suits peptides for several converging reasons. Peptides are amphipathic polymers that carry both charged and hydrophobic side chains, so they interact strongly enough with an alkyl surface to be retained, yet dissolve readily in the water and acetonitrile mixtures used to elute them. The acidic eluents that are standard in this mode hold most peptides in a protonated, well-behaved state and suppress unwanted interactions with the silica support. Resolving power is high, because retention on a hydrophobic surface integrates contributions from every residue in the sequence and is therefore sensitive to small composition changes. The same chemistry scales from analytical columns to preparative columns, so the method used to characterize a material is closely related to the method used to purify it.
For laboratory quality control, the practical appeal is that one generic method, a C18 column with an acidified acetonitrile gradient and far ultraviolet detection, gives a first-pass separation for a very wide range of synthetic sequences without compound-specific development. That generality is also the limitation. A generic gradient is not guaranteed to resolve every closely related impurity, non-peptidic components such as counterions and residual solvents are largely invisible to an ultraviolet detector, and anything that fails to elute within the run is simply absent from the record rather than flagged. Those boundaries are the reason method development, system suitability and orthogonal conditions all appear later in this brief.
How does reverse phase chromatography separate peptides at the molecular level?
Separation in reverse phase chromatography of peptides is better described as adsorption and desorption than as partitioning into a liquid-like layer, because a peptide binds to the alkyl surface through a preferred hydrophobic contact region rather than dissolving into the bonded phase. A folded or partly structured peptide presents one face to the surface, and the free energy of that contact determines how much organic solvent is needed to displace it.
That contact-region behaviour produces the steep, almost switch-like dependence of peptide retention on eluent composition that every peptide chromatographer works around. Small changes in the organic fraction of the mobile phase produce very large changes in retention, so a peptide tends to sit immobile at the column head until the local composition reaches its release point, after which it moves through the bed quickly. Two consequences follow: peaks are narrow, which helps resolution, and fixed-composition elution is impractical, which is why gradients are used.
Retention is additive to a first approximation, and the literature on peptide retention prediction is built on that approximation. Work reported in Determination of intrinsic hydrophilicity/hydrophobicity of amino acid side chains in peptides in the absence of nearest-neighbor or conformational effects (Biopolymers, 2006) used model peptides designed to isolate the contribution of a single side chain, and a later comparison reported in Intrinsic amino acid side-chain hydrophilicity/hydrophobicity coefficients determined by reversed-phase high-performance liquid chromatography of model peptides (Biopolymers, 2009) set those chromatographically derived coefficients against other hydrophobicity scales. Studies of that kind suggest that residue contributions explain much of the observed retention but not all of it, because neighbouring residues and induced secondary structure modify what the surface actually sees.
Sequence-level prediction models extend the additive picture. The algorithm described in Sequence-specific retention calculator. Algorithm for peptide retention prediction in ion-pair RP-HPLC: application to 300- and 100-A pore size C18 sorbents (Anal Chem, 2006) incorporates sequence position, peptide length and helicity terms, and reports that prediction accuracy depends on the sorbent pore size as well as on the eluent. Related work on retention calibration appears in Peptide retention standards and hydrophobicity indexes in reversed-phase high-performance liquid chromatography of peptides (Anal Chem, 2009), which describes expressing retention as a composition-based hydrophobicity index so that results transfer between gradients and instruments. The limits are clear from the studies themselves: these models were trained largely on unmodified, mostly tryptic peptides under specific acidic ion-pairing conditions, and research suggests their accuracy degrades for cyclic peptides, lipidated or otherwise modified sequences, and peptides that aggregate on the column.
Which HPLC column chemistry suits peptide analysis?
The default HPLC column chemistry for peptides is octadecylsilane, or C18, bonded to high-purity porous silica, with shorter alkyl chains such as C8 and C4 reserved for larger or strongly retained molecules. The choice is not only about chain length. Pore size, particle size and morphology, bonding density, end-capping, support purity and column dimensions all shape the chromatogram.
Chain length sets retentivity and, to a lesser degree, selectivity. C18 gives the strongest hydrophobic interaction and is appropriate for small and mid-sized peptides that would otherwise elute near the void. C8 and C4 release large, hydrophobic or highly structured molecules at lower organic content and often give better peak shape for them, because the slower mass transfer and conformational unfolding that broaden large analytes on C18 are reduced. Phenyl, biphenyl and pentafluorophenyl phases offer aromatic and dipole interactions that change the elution order relative to alkyl phases, which makes them useful as orthogonal confirmation rather than as first-choice columns.
Pore size matters because a peptide must reach the internal surface where nearly all of the bonded phase lives. Sorbents with roughly 100 angstrom pores give a large surface area and suit short peptides, while wide-pore sorbents near 300 angstrom are conventionally used for longer chains and small proteins that diffuse poorly into narrow pores. The retention prediction work cited above reported that pore size itself alters retention behaviour, not only efficiency, so a method transferred between pore sizes should not be assumed to keep its elution order.
Particle characteristics control efficiency and backpressure. Fully porous 5 micron and 3 micron particles remain common for routine purity runs, sub-2 micron particles deliver sharper peaks at the cost of pressure and stricter instrument requirements, and superficially porous particles provide much of the efficiency gain at moderate pressure. Column dimensions follow from the goal: longer beds increase peak capacity for complex sequences, narrower bores lower solvent consumption and raise mass sensitivity, and both change the gradient volume that must be recalculated during transfer.
Support quality is the quiet variable. Residual acidic silanols interact with basic side chains such as arginine, lysine and histidine, and that mixed-mode interaction is a common cause of tailing. Modern high-purity, well end-capped and hybrid particles reduce it, and the strongly acidic eluents described in the next section suppress what remains. Silica-based phases are also attacked slowly at very low pH and elevated temperature, which is why column lifetime in peptide laboratories is typically shorter than in neutral-pH assays, and why polymeric or hybrid particles are chosen when high-pH conditions are needed for selectivity.
What goes into the mobile phase, and what does trifluoroacetic acid actually do?
The classical peptide mobile phase is a pair of solvents: eluent A is water with an acid, eluent B is acetonitrile with the same acid, and trifluoroacetic acid at a low level is the most common additive. The acid performs three jobs at once, and understanding them explains most mobile phase decisions.
First, it fixes pH near 2, which protonates carboxyl groups and keeps acidic residues uncharged, so peptides behave as a consistent species rather than a mixture of ionization states that would broaden peaks. Second, it suppresses ionization of residual silanols on the support, reducing the mixed-mode tailing described above. Third, and specific to perfluorinated acids, the trifluoroacetate anion pairs with protonated basic side chains and the amino terminus, masking positive charge and making the peptide effectively more hydrophobic. Retention therefore increases and peak shape improves, and the magnitude of the increase scales with how many basic residues a sequence carries.
Swapping the ion-pairing acid is one of the most powerful selectivity tools available, because the effect is composition-dependent rather than uniform. Stronger pairing agents such as heptafluorobutyric acid push basic peptides much later and can separate pairs that co-elute with trifluoroacetic acid, while weaker additives such as formic acid or ammonium salts at low concentration reduce retention and often reduce resolution between basic variants. Phosphoric acid and phosphate buffers give excellent peak shape and low far ultraviolet absorbance but are non-volatile, so they are confined to analytical work where fraction recovery is not required.
The organic modifier is normally acetonitrile. It has low viscosity, which keeps backpressure manageable, high elution strength for peptides, and a low ultraviolet cutoff that permits detection in the far ultraviolet. Methanol is a weaker eluent with different hydrogen bonding behaviour, so it changes selectivity and sometimes resolves pairs acetonitrile cannot, but it raises viscosity and absorbs more strongly below about 220 nm, which compromises sensitive far ultraviolet detection. Isopropanol is sometimes blended in for very hydrophobic sequences that resist elution.
Practical eluent handling carries its own rules. Trifluoroacetic acid absorbs in the far ultraviolet, so an unequal acid concentration between the aqueous and organic eluents produces baseline drift that rises or falls across the gradient; laboratories commonly trim the acid level in the organic eluent slightly to flatten that drift. Both eluents should be prepared fresh in clean glassware and degassed, since dissolved gas causes pump instability and detector spikes, and acetonitrile quality matters because trace ultraviolet-absorbing contaminants concentrate on the column during equilibration and appear as ghost peaks when the gradient runs.
How is gradient elution set up for peptide HPLC?
Gradient elution peptide HPLC uses a programmed rise in the organic eluent, typically linear, from a low starting composition to a high ending composition over a defined time, because peptides release from the stationary phase over such a narrow composition window that a fixed-composition eluent would either hold them indefinitely or sweep them out unretained. The gradient converts that switch-like behaviour into an orderly, hydrophobicity-ordered separation.
Method development usually begins with a broad scouting gradient, for example 5 to 95 percent organic over 30 to 60 minutes, run to locate where the main component and anything near it elute. The elution window observed there is then used to construct a focused gradient, a shallower ramp over a narrower composition range that spreads closely eluting species apart. Gradient slope, expressed as percent organic per minute, is the dominant resolution control in this mode: shallower slopes increase peak capacity and improve the separation of near neighbours, at the cost of run time, broader peaks and lower apparent sensitivity. A slope near 1 percent per minute is a widely used compromise for analytical peptide purity work, with steeper slopes for simple separations and much shallower slopes for sequences that differ by a single residue.
What the column actually experiences is not slope in time but gradient volume relative to column volume, which is why transferring a method between column dimensions or flow rates requires rescaling rather than copying the timetable. Keeping the ratio of gradient volume to column volume constant preserves selectivity when a method moves from a 5 micron to a sub-2 micron particle or from a wide bore to a narrow bore column. Instrument dwell volume, the volume between the point of mixing and the column head, adds a delay before the gradient reaches the bed; dwell volumes differ substantially between a low-pressure mixing quaternary system and a high-pressure binary system, which is a frequent cause of shifted retention times when a method moves between instruments.
The end of the run matters as much as the ramp. A short hold at high organic content washes strongly retained material off the bed so it does not appear in the following run, and re-equilibration at the starting composition for several column volumes restores the initial state. Insufficient re-equilibration is one of the most common sources of drifting retention times across a sequence of runs.
Column temperature is the other routine variable. Raising temperature lowers eluent viscosity and backpressure, speeds mass transfer and usually sharpens peaks, and it can change selectivity by altering peptide conformation at the surface. Many peptide methods run between 30 and 60 degrees Celsius, controlled rather than ambient, because selectivity shifts with temperature and uncontrolled laboratory temperature makes retention times irreproducible. The trade is accelerated hydrolysis of silica-based phases at low pH and high temperature, which shortens column life.
Why does peptide HPLC detect at 214 nm?
Peptide HPLC detects at 214 nm because the amide bond that links every residue absorbs strongly in the far ultraviolet, roughly between 190 and 230 nm, so a detector set there responds to the peptide backbone itself rather than to side chains that may or may not be present. The practical consequence is universality: a short sequence with no tryptophan, tyrosine or phenylalanine is still seen clearly, and relative peak areas across a chromatogram remain broadly comparable.
The contrast with 280 nm makes the point. Absorbance near 280 nm comes almost entirely from aromatic side chains and, to a small extent, from disulfide bonds, which is the basis of the calculation described in Calculation of protein extinction coefficients from amino acid sequence data (Anal Biochem, 1989). That approach works well for proteins containing tryptophan and tyrosine and fails entirely for peptides that contain neither, since such a sequence is effectively transparent at 280 nm. The alternative of measuring the backbone in the far ultraviolet has a long history in protein quantitation, described in Measurement of protein by spectrophotometry at 205 nm (Anal Biochem, 1974), which reports that absorbance at that wavelength tracks peptide bond content with comparatively little dependence on composition.
Why 214 nm rather than 205 nm, where the backbone absorbs even more strongly, is a compromise with the eluent. Acetonitrile, water and especially trifluoroacetic acid all absorb increasingly as wavelength falls, so moving the detector lower raises both analyte signal and background noise and drift. Detection at 214 nm, sometimes 215 nm or 220 nm, retains most of the backbone response while keeping the baseline usable with acidified acetonitrile gradients. Laboratories that need maximum sensitivity sometimes lower the wavelength and switch to a phosphate-based or lower-absorbing acid system to make that possible.
Several caveats belong with far ultraviolet detection. Response is not identical for all species: aromatic residues, certain protecting groups and some chromophoric impurities absorb more per unit mass than a plain backbone, so an area percentage is a chromatographic number, not a mass fraction. Non-chromophoric components, including counterions, inorganic salts and water, contribute little or nothing at this wavelength. Baseline drift from acid mismatch, discussed above, can bias integration if it is not corrected. A diode array detector mitigates several of these issues by recording a full spectrum across each peak, which allows peak homogeneity checks and lets the same run be reviewed at a second wavelength.
How is chromatographic purity calculated from the chromatogram?
Chromatographic purity is normally reported as area percent, the integrated area of the main peak divided by the summed area of all integrated peaks in the run, with the solvent front and any peaks also present in a blank gradient excluded. The arithmetic is trivial; the judgement sits in the integration.
Baseline placement is the first judgement. Peaks that sit on a rising baseline, as they do in gradients with unmatched acid, need a consistent baseline convention, and the chosen convention should be recorded with the method rather than adjusted run by run. Peak threshold and minimum area settings decide which small features are counted as peaks and which are treated as noise, and changing those settings changes the reported number without changing the sample. Partially resolved pairs raise a further question, because a valley drop and a tangential skim can apportion area quite differently between a main peak and a shoulder.
A blank gradient run, with no sample loaded, is the control that makes the rest interpretable. Anything appearing in the blank originates from the eluents, the sample solvent or the system, and should not be counted against the material. Running the blank under identical conditions also reveals contamination accumulating on the column during equilibration, which otherwise shows up as reproducible ghost peaks.
The limits of the calculation deserve to be stated plainly. Area percent assumes equal detector response across all components, which is only approximately true at 214 nm. It cannot describe anything that does not absorb, so counterion content, residual solvents and water are outside its scope and require separate techniques. It cannot describe anything that fails to elute, since strongly retained or irreversibly adsorbed material simply never reaches the detector. And it cannot resolve what the gradient does not separate, so a co-eluting impurity is counted inside the main peak. For these reasons a second, deliberately different set of conditions, such as a different column chemistry, a different organic modifier or a different ion-pairing acid, is used when confirmation matters; agreement between two orthogonal separations is far stronger evidence than a single chromatogram.
Which peptide impurities does the method need to resolve?
The impurity profile that a peptide method must handle is dominated by species chemically close to the target sequence, which is exactly why a shallow gradient and careful selectivity work are necessary. Synthesis-related, storage-related and handling-related species each appear in characteristic ways.
Synthesis contributes deletion sequences, where a residue was skipped during chain assembly, and truncated sequences, where chain growth stopped. A deletion variant differs from the target by one residue and therefore by a small change in hydrophobicity, so it often elutes close to the main peak and may require a very shallow ramp to separate. Incomplete removal of side-chain protecting groups leaves adducts that usually increase hydrophobicity and elute later, while scavenger adducts and alkylation products form a second family of late-eluting species. The side reactions that generate these impurities are catalogued in the solid phase synthesis literature, including the review Solid phase peptide synthesis utilizing 9-fluorenylmethoxycarbonyl amino acids (Int J Pept Protein Res, 1990), which describes deprotection, coupling and cleavage chemistry together with the by-products each step can produce.
Chemical modification of a finished peptide produces a different set. Oxidation of methionine, cysteine or tryptophan adds polarity and usually shifts the species earlier in a reversed-phase gradient. Deamidation of asparagine or glutamine converts an amide into a carboxylate and generates isoaspartate forms, which often appear as closely spaced peaks near the parent. Epimerization during synthesis creates diastereomers that are chemically identical in composition yet differ in three-dimensional shape, and research suggests these are frequently separable in this mode because the contact surface presented to the stationary phase changes.
Higher-order species complete the picture. Disulfide-containing peptides can form scrambled isomers, dimers and oligomers, which generally elute differently from the monomer and may show broad or shouldered peaks when interconversion occurs on the timescale of the run. Aggregated material sometimes fails to elute at all and instead fouls the column head, producing declining recovery across a sequence of runs rather than a visible extra peak.
Non-peptidic components are the blind spot. Trifluoroacetate counterion, inorganic salts and residual synthesis solvents are either unretained or non-absorbing at the detection wavelength and are therefore not part of a chromatographic purity figure. Laboratories that need that information use ion chromatography, headspace gas chromatography, Karl Fischer titration or elemental methods, each of which answers a question this method does not.
How is a peptide reversed-phase method developed and optimized?
Method development proceeds from a broad scouting run to a focused, documented set of conditions, changing one variable at a time and judging each change by resolution between the main peak and its nearest neighbour rather than by overall appearance. The order in which variables are tried matters, because some change selectivity and others only change efficiency or speed.
The usual sequence begins with the scouting gradient on a C18 column with an acidified acetonitrile system, which establishes whether the sequence elutes at all, where it elutes, and how crowded the surrounding region is. Gradient slope is then reduced around the elution window, which almost always improves resolution of near neighbours. If a critical pair remains unresolved, selectivity rather than efficiency is the problem, and the effective levers are the ion-pairing acid, the organic modifier, the temperature, the pH and the column chemistry, roughly in that order of impact for peptides. Switching from trifluoroacetic to a stronger or weaker pairing acid rearranges basic peptides relative to neutral ones; switching acetonitrile for methanol or adding isopropanol alters hydrogen bonding; changing temperature alters conformation; changing from C18 to a phenyl or wide-pore C8 phase alters the contact surface.
Loading is a constraint rather than a lever. Columns overload long before they are visibly saturated, and the first sign is a peak that becomes asymmetric and fronting while resolution from an adjacent impurity collapses. Because small impurities must stay above the noise, a purity method sits between the limit set by overload and the limit set by sensitivity, and that working range should be established during development rather than assumed.
The solvent in which the sample reaches the column also matters. When the sample solvent is substantially stronger than the starting mobile phase, the band is not focused at the column head and peaks broaden, split or distort. Matching sample solvent strength to the initial gradient composition, or at least keeping it weaker, is standard practice in the chromatography literature for this reason.
Finally, development should include confirmation that the main peak is a single species as far as the method can tell. Spectral comparison across the leading edge, apex and trailing edge of the peak with a diode array detector is the routine check, and a second orthogonal separation is the stronger one. Every setting that influences the outcome, including column lot and dimensions, eluent composition, gradient table, flow rate, temperature, wavelength, equilibration time and integration parameters, belongs in the written method, since any of them can change the result.
What system suitability and validation checks show a peptide method is working?
System suitability is a short set of measurements run before and during an analytical sequence to demonstrate that the column, eluents and instrument are performing as they did when the method was established. For peptide purity work the usual parameters are resolution between the main peak and the closest specified impurity, peak asymmetry or tailing factor, apparent efficiency expressed as plate count, retention time reproducibility across replicate sample introductions, and baseline noise and drift over the gradient.
Each parameter catches a different failure. Falling resolution points to a degraded column or a change in eluent composition. Rising tailing suggests silanol activity, a partly blocked frit or a void forming at the column head. Drifting retention times point to incomplete re-equilibration, temperature instability or a pump delivering an inaccurate composition. Growing baseline noise often traces to eluent quality or to a worn detector lamp, which matters more at 214 nm than at longer wavelengths because lamp output is lower in the far ultraviolet.
A sensitivity check is included when small impurities are being counted. A sample prepared at low concentration relative to the working level is run to confirm that a peak at the reporting threshold is still detected with acceptable signal to noise. Without that check, an impurity profile can look clean simply because the system has lost sensitivity.
Formal validation goes further and is described in international frameworks such as the ICH Q2 guideline on analytical procedure validation. The elements commonly addressed are specificity, meaning that the method distinguishes the target from impurities and blanks; linearity and range over the concentrations at which the method is used; accuracy and precision, including repeatability and intermediate precision across days, analysts and instruments; detection and quantitation limits for impurity measurement; and robustness, meaning deliberate small variations in gradient slope, temperature, wavelength, acid concentration and column lot to establish which parameters must be tightly controlled. Robustness testing is particularly informative for peptide methods, because selectivity in this mode is sensitive to the ion-pairing acid level and to temperature in ways that are easy to overlook.
Which artifacts and failure modes distort peptide chromatograms?
Most disappointing peptide chromatograms come from a short list of recurring problems, and nearly all of them are diagnosable with blank runs, a column wash and fresh eluents. Recognizing the signature of each saves considerable time.
Tailing peaks on a silica-based column usually indicate interaction between basic side chains and residual silanols, or interaction with metal surfaces in the flow path for peptides bearing phosphate or carboxylate clusters. Stronger acid, a higher-purity or hybrid particle, or an inert flow path addresses it. Fronting or splitting at the apex points instead to overload, to a sample solvent stronger than the starting eluent, or to a physical void at the column inlet after prolonged use at low pH.
Ghost peaks that appear in blank gradients arise from contaminants in the aqueous eluent that concentrate at the column head during equilibration and then elute when organic content rises. Peaks that appear in a run but belong to the previous sample are carryover, usually from the sample path or from strongly retained material released by the next gradient, and a high-organic wash step at the end of each run reduces both.
Baseline drift across the gradient is almost always an acid mismatch between the aqueous and organic eluents, since trifluoroacetic acid absorbs in the far ultraviolet. Irregular baseline spikes suggest bubbles from inadequately degassed eluent. Retention times that creep in one direction across a sequence suggest incomplete re-equilibration, slow temperature change or evaporation of the more volatile organic component from an open reservoir.
Low or declining recovery is the subtlest failure, because nothing on the chromatogram looks wrong. Hydrophobic and aggregation-prone peptides can adsorb to surfaces in the flow path or remain bound to the stationary phase, so the peak observed represents less material than was loaded. The literature on hydrophobic peptide handling describes this behaviour and the use of stronger organic washes and alternative surface chemistries to reduce it. Finally, silica-based columns used at low pH and elevated temperature lose bonded phase over time, which shows as gradually shorter retention and poorer resolution; tracking system suitability results across a column lifetime makes that decline visible before it invalidates a run.
How is the method documented, and where does independent testing fit?
A peptide purity analysis is only as transferable as its documentation, so the record should contain the column identity and dimensions, eluent composition including the acid and its level, the complete gradient table with flow rate and equilibration time, column temperature, detection wavelength and bandwidth, sample solvent, system suitability criteria and results, the integration parameters used, and the chromatogram itself with the baseline shown. A reported figure without those conditions cannot be reproduced or compared, because two laboratories running different gradients, columns and integration settings can report different numbers for the same material without either being careless.
Independent testing adds the separation of interest from the result. Analysis performed by a laboratory with no stake in the outcome, under documented conditions, produces a record that can be examined rather than taken on trust. Real Peptides supplies research-use-only compounds through its catalog and publishes independent third-party certificates of analysis for every batch at /coas/, so the analytical record associated with a given lot is publicly available rather than summarized. Researchers evaluating any supplied compound can also confirm its CAS number and molecular identity against the published chemistry before procurement, and company and service information is listed at /location/.
Real Peptides does not provide preparation or use guidance for the compounds it supplies, because they are research-use-only materials intended for laboratory work under the direction of a qualified investigator. Where quantity needs to be expressed at all, the appropriate framework is concentration, mass of compound per volume of solvent in milligrams per millilitre, a laboratory unit that describes a solution rather than any use of it. Analytical questions, including the chromatographic conditions described throughout this brief, belong to the receiving laboratory and its own validated methods.
All compounds discussed on this page are sold for laboratory research use only and are not for human consumption.
References
- Measurement of protein by spectrophotometry at 205 nm. Anal Biochem. 1974. PMID: 4407487
- Calculation of protein extinction coefficients from amino acid sequence data. Anal Biochem. 1989. PMID: 2610349
- Solid phase peptide synthesis utilizing 9-fluorenylmethoxycarbonyl amino acids. Int J Pept Protein Res. 1990. PMID: 2191922
- Determination of intrinsic hydrophilicity/hydrophobicity of amino acid side chains in peptides in the absence of nearest-neighbor or conformational effects. Biopolymers. 2006. PMID: 16315143
- Sequence-specific retention calculator. Algorithm for peptide retention prediction in ion-pair RP-HPLC: application to 300- and 100-A pore size C18 sorbents. Anal Chem. 2006. PMID: 17105172
- Intrinsic amino acid side-chain hydrophilicity/hydrophobicity coefficients determined by reversed-phase high-performance liquid chromatography of model peptides: comparison with other hydrophilicity/hydrophobicity scales. Biopolymers. 2009. PMID: 19795449
- Peptide retention standards and hydrophobicity indexes in reversed-phase high-performance liquid chromatography of peptides. Anal Chem. 2009. PMID: 19848410